Innovations in Communications™ Since 1955

LEE DAN® COMMUNICATIONS, INC.
155 Adams Avenue, Hauppauge, New York, 11788-3699 USA
Phone: 800-231-1414 or 631-231-1414 Fax: 631-231-1498 E-Mail: info@leedan.com
Loading
                              
IR-319B
(IR319B)
IR-320B
(IR320B)
Single & Dual Patient Stations


The IR319B (IR-319B) Single Patient Station provides reliable two-way, hands-free communication between patient and nurse, and provides for programmable privacy, paging and patient priority from the master station. Calling the nurse is accomplished by pressing the call cord button or the area indicated NURSE on the pillow speaker. Solid state LEDs on the panel indicate the call has been placed and the intercom is in use.

If the pillow speaker or call cord is accidentally removed from its receptacle, the nurse master station registers a noncancellable cord out call. The call may be reset only by returning the call cord or pillow speaker to its receptacle.        

The IR320B (IR-320B) Dual Patient Station has the same specifications and functions as the IR319B, plus an additional independently operating call cord/pillow speaker receptacle and a second call-placed indicator.

FEATURES

UL® 1069 Listed

Supervised

All stations interchangeable

Input available for Hill-Room® Sidecome® beds or remote call-in button applications

Uses pillow speakers or push-button call cords

SPECIFICATIONS  

 Dimensions: 4.5" (114 mm) H, 6.375" (162mm) W, 2.75" (70mm)D.  Projects: 1" (25.4 mm) above mounting surface

Finish: Light texture, white

Speaker: 2.5" x 2.5" (64 mm x 64 mm)

Connections: Plug-in type

Housing: Steel City #H3BD and #3GC plaster ring, or exact equal. Horizontal mount.

 

 

REQUIRED COMPONENTS

Tek-CARE® NC300II Nurse Call System

 

 
Click for Spec Sheet

Architects' & Engineers' Specifications:

Patient Stations shall be TekTone®'s IR319B (IR-319B), IR320B (IR-320B), or approved equal as supplied by LEE DAN®. Stations, and wiring from the Central Equipment (CE) to stations, shall be supervised. The patient stations shall include solid-state LSI/microprocessor technology, yellow call-placed indicator(s), and combination red in-use indicator and reset button. Single patient stations shall include one 8-conductor modular DIN-style jack receptacle for pillow speakers or patient's call cord. Dual patient stations shall include two jacks.

The station shall use a 2.5"x2.5" speaker/microphone for accurate voice reproduction. Removal of the call cord or pillow speaker from the receptacle shall result in placing a call with the indication of cord out at the master station, which may be reset only by replacing the call cord or pillow speaker in its receptacle. All station components shall be mounted on an attractive, flame-retardant (UL® 94V-0) plastic panel. Connections shall be plug-in type.

The IR319B, IR320B shall be UL® 1069 Listed.

| Brochure | Doc | Spec Sheet | User Guide | Manual | Wiring |